A Comparison of Readings from Thermoluminescent Dosimeter Ring Badges Worn in Different Positions

dc.contributor.authorOsborne, Keexia L.
dc.contributor.authorNielsen, Cybil J.
dc.contributor.authorBullock, John R.
dc.contributor.authorJennings, S. Gregory
dc.contributor.departmentRadiology and Imaging Sciences, School of Medicineen_US
dc.date.accessioned2022-03-25T19:37:04Z
dc.date.available2022-03-25T19:37:04Z
dc.date.issued2020-12
dc.description.abstractOur purpose was to evaluate whether the position of a thermoluminescent dosimeter (TLD) crystal results in different exposure readings. Methods: Nine subjects wore 2 TLD badges (one facing inward, toward the palm, and one facing outward) for 2 mo. Both TLDs were worn on the middle finger of the dominant hand, with the inward-facing TLD placed at the bottom and the outward-facing TLD at the top. At the end of the first month, these TLDs were replaced with new ones for another month. Combined results from the badges for the 2 mo were recorded in millisieverts. A paired t test with 2-sample means was performed to compare the 2 positions in general nuclear medicine and PET/CT subjects, with an α of 0.05. Results: For all subjects and for the general nuclear medicine and PET/CT groups, mean exposure was greater for the inward-facing TLD. Conclusion: For a TLD worn on the dominant hand, extremity-exposure readings are maximized when the TLD faces inward.en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.citationOsborne, K. L., Nielsen, C. J., Bullock, J. R., & Jennings, S. G. (2020). A Comparison of Readings from Thermoluminescent Dosimeter Ring Badges Worn in Different Positions. Journal of Nuclear Medicine Technology, 48(4), 361–362. https://doi.org/10.2967/jnmt.120.243345en_US
dc.identifier.urihttps://hdl.handle.net/1805/28312
dc.language.isoenen_US
dc.publisherSNMMIen_US
dc.relation.isversionof10.2967/jnmt.120.243345en_US
dc.relation.journalJournal of Nuclear Medicine Technologyen_US
dc.rightsPublisher Policyen_US
dc.sourcePublisheren_US
dc.subjectcrystalen_US
dc.subjectpositionen_US
dc.subjectthermoluminescent dosimeteren_US
dc.titleA Comparison of Readings from Thermoluminescent Dosimeter Ring Badges Worn in Different Positionsen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Osborne2020Comparison-PP.pdf
Size:
451.45 KB
Format:
Adobe Portable Document Format
Description:
License bundle