In Vitro Performance of Near Infrared Light Transillumination at 780-nm and Digital Radiography for Detection of Non-Cavitated Approximal Caries

dc.contributor.authorAbogazalah, Naif
dc.contributor.authorEckert, George J.
dc.contributor.authorAndo, Masatoshi
dc.contributor.departmentDepartment of Cariology, Operative Dentistry and Dental Public Health, School of Dentistryen_US
dc.date.accessioned2017-06-30T16:59:51Z
dc.date.available2017-06-30T16:59:51Z
dc.date.issued2017
dc.description.abstractObjectives To evaluate the ability of a Near Infrared Light Transillumination (NILT) device to detect non-cavitated approximal caries lesions; and to compare its performance to Digital Radiography (DR). Methods Thirty human extracted premolars (sound to lesions into the outer one-third of dentin) were selected. Lesion depth was confirmed by micro-computed tomography (μ-CT). Teeth were mounted in a custom-made device to simulate approximal contact. DR and NILT (CariVu™, DEXIS, LLC, Hatfield, PA, USA) examinations were performed and repeated by three trained and calibrated examiners. Sensitivity, specificity, area under ROC curve (Az), inter- and intra-class correlation coefficients (ICCs) for each method, and correlation among the methods were determined. Results ICCs for intra-/inter-examiner agreement were substantial for NILT (0.69/0.64), and moderate for DR (0.52/0.48). Sensitivity/specificity for NILT and DR were 0.68/0.93 and 0.50/0.64, respectively. Az for NILT was 0.81, while for DR it was 0.61. Spearman correlation coefficient with μ-CT for NILT (0.65, p<0.001) demonstrated moderate association, while that of DR suggested no association (0.19, p = 0.289). Conclusion Within the limitations of this in vitro study, NILT demonstrated a potential for early approximal caries detection. NILT and DR performed the same regarding the accuracy for non-cavitated approximal caries detection; however, NILT was superior to DR in terms of repeatability, agreement and correlation with μ-CT.en_US
dc.eprint.versionAuthor's manuscripten_US
dc.identifier.citationAbogazalah, N., Eckert, G. J., & Ando, M. (2017). In Vitro Performance of Near Infrared Light Transillumination at 780-nm and Digital Radiography for Detection of Non-Cavitated Approximal Caries. Journal of Dentistry. https://doi.org/10.1016/j.jdent.2017.05.018en_US
dc.identifier.urihttps://hdl.handle.net/1805/13298
dc.language.isoenen_US
dc.publisherElsevieren_US
dc.relation.isversionof10.1016/j.jdent.2017.05.018en_US
dc.relation.journalJournal of Dentistryen_US
dc.rightsPublisher Policyen_US
dc.sourceAuthoren_US
dc.subjectapproximal cariesen_US
dc.subjectcaries detectionen_US
dc.subjectnear infrared transilluminationen_US
dc.titleIn Vitro Performance of Near Infrared Light Transillumination at 780-nm and Digital Radiography for Detection of Non-Cavitated Approximal Cariesen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Abogazalah_2017_invitro.pdf
Size:
607.05 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.88 KB
Format:
Item-specific license agreed upon to submission
Description: