In Vitro Performance of Near Infrared Light Transillumination at 780-nm and Digital Radiography for Detection of Non-Cavitated Approximal Caries
Date
Language
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Abstract
Objectives
To evaluate the ability of a Near Infrared Light Transillumination (NILT) device to detect non-cavitated approximal caries lesions; and to compare its performance to Digital Radiography (DR).
Methods
Thirty human extracted premolars (sound to lesions into the outer one-third of dentin) were selected. Lesion depth was confirmed by micro-computed tomography (μ-CT). Teeth were mounted in a custom-made device to simulate approximal contact. DR and NILT (CariVu™, DEXIS, LLC, Hatfield, PA, USA) examinations were performed and repeated by three trained and calibrated examiners. Sensitivity, specificity, area under ROC curve (Az), inter- and intra-class correlation coefficients (ICCs) for each method, and correlation among the methods were determined.
Results
ICCs for intra-/inter-examiner agreement were substantial for NILT (0.69/0.64), and moderate for DR (0.52/0.48). Sensitivity/specificity for NILT and DR were 0.68/0.93 and 0.50/0.64, respectively. Az for NILT was 0.81, while for DR it was 0.61. Spearman correlation coefficient with μ-CT for NILT (0.65, p<0.001) demonstrated moderate association, while that of DR suggested no association (0.19, p = 0.289).
Conclusion
Within the limitations of this in vitro study, NILT demonstrated a potential for early approximal caries detection. NILT and DR performed the same regarding the accuracy for non-cavitated approximal caries detection; however, NILT was superior to DR in terms of repeatability, agreement and correlation with μ-CT.