Abstract

PURPOSE OF REVIEW:

Sacral neuromodulation (SNM) is being used to treat lower urinary tract symptoms (LUTS) with growing popularity among clinicians in multiple specialties. As this therapy becomes more common in the USA and Europe, urologists will encounter more patients implanted with SNM generators. RECENT FINDINGS:

Over time, it has recently been understood that up to 53% will develop pain at the implant site as reported by Groen et al. (J Urol 186:954, 2011) and 3-38% will lose effective stimulation as reported by Al-zahrani et al. (J Urol 185:981, 2011) and White et al. (Urology 73:731, 2009). There is a paucity of troubleshooting methodology in the literature, apart from revision surgery, to salvage the SNM generator. In fact, it has been suggested that one contemporary series' failure rate is lower than some historic series because of the ability to reprogram devices as reported by Siegel et al. (J Urol 199:229, 2018). Standard algorithms for such reprogramming efforts are lacking in the literature and may salvage some patients otherwise destined for surgical revision or addition of multimodal therapy to achieve acceptable symptom control. It is possible to troubleshoot and thereby salvage many SNM generators, saving patients from surgical revision in many cases and increasing the number of patients with persistent benefit from SNM. The algorithms presented in this manuscript represent a systematic strategy for reprogramming and troubleshooting SNM generators.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Powell C. R. (2018). Troubleshooting Interstim Sacral Neuromodulation Generators to Recover Function. Current urology reports, 19(10), 86. doi:10.1007/s11934-018-0837-5
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Current Urology reports
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Full Text Available at
This item is under embargo {{howLong}}