Motion artifacts in kidney stone imaging using single-source and dual-source dual-energy CT scanners: a phantom study

Date
2015-10
Language
American English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Springer
Abstract

PURPOSE:

Dual-energy computed tomography (DECT) has shown the capability of differentiating uric acid (UA) from non-UA stones with 90-100% accuracy. With the invention of dual-source (DS) scanners, both low- and high-energy images are acquired simultaneously. However, DECT can also be performed by sequential acquisition of both images on single-source (SS) scanners. The objective of this study is to investigate the effects of motion artifacts on stone classification using both SS-DECT and DS-DECT. METHODS:

114 kidney stones of different types and sizes were imaged on both DS-DECT and SS-DECT scanners with tube voltages of 80 and 140 kVp with and without induced motion. Postprocessing was conducted to create material-specific images from corresponding low- and high-energy images. The dual-energy ratio (DER) and stone material were determined and compared among different scans. RESULTS:

For the motionless scans, all stones were correctly classified with SS-DECT, while two cystine stones were misclassified with DS-DECT. When motion was induced, 94% of the stones were misclassified with SS-DECT versus 11% with DS-DECT (P < 0.0001). Stone size was not a factor in stone misclassification under motion. Stone type was not a factor in stone misclassification under motion with SS-DECT, although with DS-DECT, cystine showed higher number of stone misclassification. CONCLUSIONS:

Motion artifacts could result in stone misclassification in DECT. This effect is more pronounced in SS-DECT versus DS-DECT, especially if stones of different types lie in close proximity to each other. Further, possible misinterpretation of the number of stones (i.e., missing one, or thinking that there are two) in DS-DECT could be a potentially significant problem.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Ibrahim, E.-S. H., Cernigliaro, J. G., Pooley, R. A., Williams, J. C., & Haley, W. E. (2015). Motion Artifacts in Kidney Stone Imaging Using Single-Source and Dual-Source Dual-Energy CT Scanners. A Phantom Study. Abdominal Imaging, 40(8), 3161–3167. http://doi.org/10.1007/s00261-015-0530-9
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Abdominal Imaging
Rights
Publisher Policy
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}