Effect of an Aerosol Box on Intubation in Simulated Emergency Department Airways: A Randomized Crossover Study

dc.contributor.authorTurner, Joseph S.
dc.contributor.authorFalvo, Lauren E.
dc.contributor.authorAhmed, Rami A.
dc.contributor.authorEllender, Timothy J.
dc.contributor.authorCorson-Knowles, Dan
dc.contributor.authorBona, Anna M.
dc.contributor.authorSarmiento, Elisa J.
dc.contributor.authorCooper, Dylan D.
dc.contributor.departmentEmergency Medicine, School of Medicineen_US
dc.date.accessioned2021-04-14T18:09:56Z
dc.date.available2021-04-14T18:09:56Z
dc.date.issued2020-11
dc.description.abstractIntroduction: The use of transparent plastic aerosol boxes as protective barriers during endotracheal intubation has been advocated during the severe acute respiratory syndrome coronavirus 2 pandemic. There is evidence of worldwide distribution of such devices, but some experts have warned of possible negative impacts of their use. The objective of this study was to measure the effect of an aerosol box on intubation performance across a variety of simulated difficult airway scenarios in the emergency department. Methods: This was a randomized, crossover design study. Participants were randomized to intubate one of five airway scenarios with and without an aerosol box in place, with randomization of intubation sequence. The primary outcome was time to intubation. Secondary outcomes included number of intubation attempts, Cormack-Lehane view, percent of glottic opening, and resident physician perception of intubation difficulty. Results: Forty-eight residents performed 96 intubations. Time to intubation was significantly longer with box use than without (mean 17 seconds [range 6-68 seconds] vs mean 10 seconds [range 5-40 seconds], p <0.001). Participants perceived intubation as being significantly more difficult with the aerosol box. There were no significant differences in the number of attempts or quality of view obtained. Conclusion: Use of an aerosol box during difficult endotracheal intubation increases the time to intubation and perceived difficulty across a range of simulated ED patients.en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.citationTurner, J. S., Falvo, L. E., Ahmed, R. A., Ellender, T. J., Corson-Knowles, D., Bona, A. M., ... & Cooper, D. D. (2020). Effect of an Aerosol Box on Intubation in Simulated Emergency Department Airways: A Randomized Crossover Study. Western Journal of Emergency Medicine, 21(6), 78. https://doi.org/10.5811/westjem.2020.8.48901en_US
dc.identifier.urihttps://hdl.handle.net/1805/25639
dc.language.isoenen_US
dc.publisherUniversity of Californiaen_US
dc.relation.isversionof10.5811/westjem.2020.8.48901en_US
dc.relation.journalWestern Journal of Emergency Medicineen_US
dc.rightsAttribution 4.0 International*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/*
dc.sourcePublisheren_US
dc.subjectSARS-CoV-2en_US
dc.subjectaerosol boxen_US
dc.subjectintubationen_US
dc.titleEffect of an Aerosol Box on Intubation in Simulated Emergency Department Airways: A Randomized Crossover Studyen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Turner2020Effect.pdf
Size:
378.23 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: