Dueling Electrospray Implemented on a Traveling-Wave Ion Mobility/Time-Of-Flight Mass Spectrometer: Towards a Gas-Phase Workbench for Structural Biology

dc.contributor.authorWebb, Ian
dc.contributor.authorMorrison, Lindsay
dc.contributor.authorBrown, Jeff
dc.date.accessioned2019-07-07T18:23:43Z
dc.date.available2019-07-07T18:23:43Z
dc.date.issued2019-07-05
dc.description• Covalent chemistry was performed in an ion mobility/mass spectrometer (IM/MS). • The effects of various parameters on the extent of reactions was observed. • Angiotensin I and ubiquitin were covalently labeled in the gas-phase. • Ion/ion reactions modify ions with solution-like conformations.en_US
dc.description.abstractThe traveling wave trap cell of a commercial ion mobility mass spectrometer (IM/MS) was used as a gas-phase reactor for covalent chemistry by making a simple modification to a standard nanoelectrospray source. Reagents and analytes were generated from pulsed opposite polarity nanoelectrospray sources and isolated by their m/z prior to reaction. Covalent bond formation was first observed with the model peptide angiotensin I. The modification site was identified as the N-terminus of the peptide by collision induced dissociation (CID). The IM cell separated the covalent reaction product from the proton transfer product by their respective ion mobilities. Next, the effects of several trapping parameters, including the trap traveling wave height, the trap RF voltage, and the trap pressure, were evaluated. Decreasing traveling wave height and increasing RF voltage and pressure increased the number of proton transfer events from apomyoglobin to reagent anions. The 6+ charge state of ubiquitin generated from nanospray under native-like conditions was covalently modified in the gas phase through ion/ion reactions. Probing the reacted protein with CID led to the assignment of lysine 29 and arginine 54 as reactive nucleophiles accessible to the reagent. IM analysis of the unmodified native-like 6+ charge state revealed that the gas-phase structure of the protein in the trap was in its compact form. Overall, we introduce a promising method for three-dimensional structural characterization of biomacromolecules.en_US
dc.identifier.citationI.K. Webb, L.J. Morrison, J. Brown, Dueling electrospray implemented on a traveling-wave ion mobility/time-of-flight mass spectrometer: Towards a gas-phase workbench for structural biology, International Journal of Mass Spectrometry (2019), doi: https://doi.org/10.1016/j.ijms.2019.116177.en_US
dc.identifier.urihttps://hdl.handle.net/1805/19851
dc.language.isoen_USen_US
dc.publisherElsevieren_US
dc.relation.isversionof10.1016/j.ijms.2019.116177.
dc.subjectNative mass spectrometryen_US
dc.subjectIon/Ion Reactionsen_US
dc.subjectCovalent Labelingen_US
dc.subjectIon mobilityen_US
dc.titleDueling Electrospray Implemented on a Traveling-Wave Ion Mobility/Time-Of-Flight Mass Spectrometer: Towards a Gas-Phase Workbench for Structural Biologyen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Webb-IJMS-2019.pdf
Size:
3.47 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: