Load System of Segmental T-Loops for Canine Retraction

dc.contributor.authorXia, Zeyang
dc.contributor.authorChen, Jie
dc.contributor.authorJiang, Feifei
dc.contributor.authorLi, Shuning
dc.contributor.authorViecilli, Rodrigo F.
dc.contributor.authorLiu, Sean Y.
dc.contributor.departmentOrthodontics and Oral Facial Genetics, School of Dentistry
dc.date.accessioned2025-04-30T13:58:43Z
dc.date.available2025-04-30T13:58:43Z
dc.date.issued2013
dc.description.abstractIntroduction: The orthodontic load system, especially ideal moment-to-force ratios, is the commonly used design parameter of segmental T-loops for canine retraction. However, the load system, including moment-to-force ratios, can be affected by the changes in canine angulations and interbracket distances. We hypothesized that clinical changes in canine position and angulation during canine retraction will significantly affect the load system delivered to the tooth. Methods: The load systems of 2 T-loop groups, one for translation and the other for controlled tipping, from 9 bilateral canine retraction patients were made to the targeted values obtained from finite element analyses and validated. Each loop was tested on the corresponding maxillary dental cast obtained in the clinic. The casts were made before and after each treatment interval so that both initial and residual load systems could be obtained. The pretreatment and posttreatment interbracket distances were recorded for calculating interbracket distance changes. Results: As the interbracket distances decreased, the average retraction-force drop per interbracket distance reduction was 36 cN/mm, a 30% drop per 1 mm of interbracket distance decrease. The average antitipping-moment drops per interbracket distance reductions were 0.02 N-mm per millimeter for controlled tipping and 1.4 N-mm per millimeter for translation, about 0.6% and 17% drops per 1 mm of interbracket decrease, respectively. Consequently, the average moment-to-force ratio increases per 1 mm of interbracket distance reduction were 1.24 mm per millimeter for controlled tipping and 6.34 mm per millimeter for translation. There was a significant residual load, which could continue to move the tooth if the patient missed the next-scheduled appointment. Conclusions: Clinical changes in canine position and angulation during canine retraction significantly affect the load system. The initial planned moment-to-force ratio needs to be lower to reach the expected average ideal value. Patients should be required to follow the office visit schedule closely to prevent negative effects because of significant moment-to-force ratios increases with time.
dc.eprint.versionAuthor's manuscript
dc.identifier.citationXia Z, Chen J, Jiangc F, Li S, Viecilli RF, Liu SY. Load system of segmental T-loops for canine retraction. Am J Orthod Dentofacial Orthop. 2013;144(4):548-556. doi:10.1016/j.ajodo.2013.05.007
dc.identifier.urihttps://hdl.handle.net/1805/47557
dc.language.isoen_US
dc.publisherElsevier
dc.relation.isversionof10.1016/j.ajodo.2013.05.007
dc.relation.journalAmerican Journal of Orthodontics and Dentofacial Orthopedics
dc.rightsPublisher Policy
dc.sourcePMC
dc.subjectDental stress analysis
dc.subjectCuspid
dc.subjectOrthodontic wires
dc.subjectOrthodontic appliance design
dc.titleLoad System of Segmental T-Loops for Canine Retraction
dc.typeArticle
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Xia2013Load-AAM.pdf
Size:
1.6 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2.04 KB
Format:
Item-specific license agreed upon to submission
Description: