Effect of HA-coating and HF etching on experemental zirconia implant evaluation using in vivo rabbit model

dc.contributor.advisorChu, Tien-Min Gabriel
dc.contributor.authorHuang, Sung-Enen_US
dc.contributor.otherJohn, Vanchit (Vanchit Kurien), 1965-
dc.contributor.otherKowolik, Michael J.
dc.contributor.otherZunt, Susan L., 1951-
dc.contributor.otherBlanchard, Steven B.
dc.date.accessioned2011-01-14T13:24:59Z
dc.date.available2011-01-14T13:24:59Z
dc.date.issued2010
dc.degree.date2010en_US
dc.degree.disciplineSchool of Dentistryen_US
dc.degree.grantorIndiana Universityen_US
dc.degree.levelM.S.D.en_US
dc.descriptionIndiana University-Purdue University Indianapolis (IUPUI)en_US
dc.description.abstractThe objective of this study was to evaluate the in vivo performance of the hydroxyapatite (HA) coating and hydrofluoric acid (HF) etching zirconia (ZrO) implants and to compare the result with titanium (Ti) implants treated in a similar manner. A total of four different implant types were tested in this study. Threaded zirconia implants with HA coating (Test 1) and zirconia implants with HF-treated surfaces (Test 2) were used to compare to the same size of titanium implants treated in identical fashion (control 1 and control 2). All implants measured about 3.5 mm at the thread diameter and 7.0 mm in total length. Each rabbit received two zirconia and two titanium implants treated in the same manner (either HA-coated or HF-etched). The samples were implanted into the rabbit tibias and retrieved at 6 weeks. Upon retrieval, 24 specimens (6 samples for each group) were fixed and dehydrated. The samples were then embedded undecalcified in PMMA for histomorphometry to quantify the bone-to-implant contact (BIC). Another 24 samples were kept in 0.9% saline and were evaluated using removal torque (RT) analysis to assess the strength of the implant-to-bone interface. The histomorphometric examination demonstrated direct bone-to-implant contact for all four groups. HA particle separation from the implants surface was seen in a majority of the HA-coated samples. No signs of inflammation or foreign body reaction were found during examination. Due to the HA particle smear contamination in the ZrO-HA group, no data was collected in this group. The mean BIC at the first three threads of the Ti-HA, Ti-HF and ZrO-HF were 57.78±18.22%, 46.41±14.55% and 47.41±14.05%, respectively. No statistically significant difference was found pair-wise among these three groups. When comparing the BIC data with the machined-surface implants, a statistically significant difference was found between the Ti-HA versus Ti implant group and the Ti-HF versus Ti implant group. The mean bone area (BA) at the first three threads for Ti-HA, Ti-HF and ZrO-HF showed statistically significant difference (p<0.05) between the ZrO-HF and Ti-HA groups, favoring the ZrO-HF group. The value of the peak removal force could only be collected from the Ti-HA group during the removal torque test. The mean RT value for the Ti-HA group was 24.39±2.58 Ncm. When comparing the RT result with our pilot study using machined-surface implants, the Ti-HA group showed statistically significant (p<0.05) higher values than the machined-surface Ti implants. The result of this study proves the in vivo biocompatibility of all four implant types tested. In the three measurable implant groups, the histomorphologic analysis showed comparable osseointegration properties in this animal model.en_US
dc.identifier.urihttps://hdl.handle.net/1805/2366
dc.identifier.urihttp://dx.doi.org/10.7912/C2/1671
dc.language.isoen_USen_US
dc.subjecthydroxyapatiteen_US
dc.subjecthydrofluoric aciden_US
dc.subjectzirconiaen_US
dc.subjectdental implanten_US
dc.subject.meshZirconium -- chemistryen_US
dc.subject.meshTitanium -- chemistryen_US
dc.subject.meshHydroxyapatites -- chemistryen_US
dc.subject.meshHydrofluoric Acid -- chemistryen_US
dc.subject.meshCoated Materials, Biocompatible -- chemistryen_US
dc.subject.meshDental Implantsen_US
dc.subject.meshOsseointegrationen_US
dc.titleEffect of HA-coating and HF etching on experemental zirconia implant evaluation using in vivo rabbit modelen_US
dc.typeThesisen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Final 3.pdf
Size:
1.72 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.88 KB
Format:
Item-specific license agreed upon to submission
Description: