Effect of HA-coating and HF etching on experemental zirconia implant evaluation using in vivo rabbit model

If you need an accessible version of this item, please email your request to digschol@iu.edu so that they may create one and provide it to you.
Date
2010
Language
American English
Embargo Lift Date
Department
Committee Chair
Degree
M.S.D.
Degree Year
2010
Department
School of Dentistry
Grantor
Indiana University
Journal Title
Journal ISSN
Volume Title
Found At
Abstract

The objective of this study was to evaluate the in vivo performance of the hydroxyapatite (HA) coating and hydrofluoric acid (HF) etching zirconia (ZrO) implants and to compare the result with titanium (Ti) implants treated in a similar manner. A total of four different implant types were tested in this study. Threaded zirconia implants with HA coating (Test 1) and zirconia implants with HF-treated surfaces (Test 2) were used to compare to the same size of titanium implants treated in identical fashion (control 1 and control 2). All implants measured about 3.5 mm at the thread diameter and 7.0 mm in total length. Each rabbit received two zirconia and two titanium implants treated in the same manner (either HA-coated or HF-etched). The samples were implanted into the rabbit tibias and retrieved at 6 weeks. Upon retrieval, 24 specimens (6 samples for each group) were fixed and dehydrated. The samples were then embedded undecalcified in PMMA for histomorphometry to quantify the bone-to-implant contact (BIC). Another 24 samples were kept in 0.9% saline and were evaluated using removal torque (RT) analysis to assess the strength of the implant-to-bone interface. The histomorphometric examination demonstrated direct bone-to-implant contact for all four groups. HA particle separation from the implants surface was seen in a majority of the HA-coated samples. No signs of inflammation or foreign body reaction were found during examination. Due to the HA particle smear contamination in the ZrO-HA group, no data was collected in this group. The mean BIC at the first three threads of the Ti-HA, Ti-HF and ZrO-HF were 57.78±18.22%, 46.41±14.55% and 47.41±14.05%, respectively. No statistically significant difference was found pair-wise among these three groups. When comparing the BIC data with the machined-surface implants, a statistically significant difference was found between the Ti-HA versus Ti implant group and the Ti-HF versus Ti implant group. The mean bone area (BA) at the first three threads for Ti-HA, Ti-HF and ZrO-HF showed statistically significant difference (p<0.05) between the ZrO-HF and Ti-HA groups, favoring the ZrO-HF group. The value of the peak removal force could only be collected from the Ti-HA group during the removal torque test. The mean RT value for the Ti-HA group was 24.39±2.58 Ncm. When comparing the RT result with our pilot study using machined-surface implants, the Ti-HA group showed statistically significant (p<0.05) higher values than the machined-surface Ti implants. The result of this study proves the in vivo biocompatibility of all four implant types tested. In the three measurable implant groups, the histomorphologic analysis showed comparable osseointegration properties in this animal model.

Description
Indiana University-Purdue University Indianapolis (IUPUI)
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Source
Alternative Title
Type
Thesis
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Full Text Available at
This item is under embargo {{howLong}}