A Handheld Quantifiable Soft Tissue Manipulation Device for Tracking Real-Time Dispersive Force-Motion Patterns to Characterize Manual Therapy Treatment

dc.contributor.authorBhattacharjee, Abhinaba
dc.contributor.authorAnwar, Sohel
dc.contributor.authorChien, Stanley
dc.contributor.authorLoghmani, M. Terry
dc.contributor.departmentPhysical Therapy, School of Health and Human Sciences
dc.date.accessioned2024-01-03T13:33:08Z
dc.date.available2024-01-03T13:33:08Z
dc.date.issued2023
dc.description.abstractObjective: Low back pain (LBP) is one of the leading neuromusculoskeletal (NMSK) problems around the globe. Soft Tissue Manipulation (STM) is a force-based, non-invasive intervention used to clinically address NMSK pain conditions. Current STM practice standards are mostly subjective, suggesting an urgent need for quantitative metrics. This research aims at developing a handheld, portable smart medical device for tracking real-time dispersive force-motions to characterize manual therapy treatments as Quantifiable Soft Tissue Manipulation (QSTM). Methods: The device includes two 3D load-cells to quantify compressive and planar-shear forces, coupled with a 6 degrees-of-freedom IMU sensor for acquiring volitionally adapted therapeutic motions while scanning and mobilizing myofascial restrictions over larger areas of the body. These force-motions characterize QSTM with treatment parameters (targeted force, application angle, rate, direction, motion pattern, time) as a part of post-processing on a PC software (Q-Ware©). A human case study was conducted to treat LBP as proof-of-concept for the device's clinical usability. Results: External validation of treatment parameters reported adequate device precision required for clinical use. The case study findings revealed identifiable therapeutic force-motion patterns within treatments indicating subject's elevated force-endurance with self-reported pain reduction. Conclusion: QSTM metrics may enable study of STM dosing for optimized pain reduction and functional outcomes using documentable manual therapy. Clinical trials will further determine its reliability and comparison to conventional STM. Significance: This medical device technology not only advances the state-of-the-art manual therapy with precision rehabilitation but also augments practice with reproducibility to examine neurobiological responses of individualized STM prescriptions for NMSK pathology.
dc.eprint.versionAuthor's manuscript
dc.identifier.citationBhattacharjee A, Anwar S, Chien S, Loghmani MT. A Handheld Quantifiable Soft Tissue Manipulation Device for Tracking Real-Time Dispersive Force-Motion Patterns to Characterize Manual Therapy Treatment. IEEE Trans Biomed Eng. 2023;70(5):1553-1564. doi:10.1109/TBME.2022.3222124
dc.identifier.urihttps://hdl.handle.net/1805/37572
dc.language.isoen_US
dc.publisherIEEE
dc.relation.isversionof10.1109/TBME.2022.3222124
dc.relation.journalIEEE Transactions on Biomedical Engineering
dc.rightsPublisher Policy
dc.sourcePMC
dc.subjectNeuromusculoskeletal disorders
dc.subjectSoft tissue manipulation
dc.subjectMedical device
dc.subjectForce-motion patterns
dc.subjectTreatment parameters
dc.subjectMassage
dc.subjectPhysical therapy
dc.subjectPrecision rehabilitation
dc.titleA Handheld Quantifiable Soft Tissue Manipulation Device for Tracking Real-Time Dispersive Force-Motion Patterns to Characterize Manual Therapy Treatment
dc.typeArticle
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
nihms-1894258.pdf
Size:
2.44 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: