Comprehensive Costs Associated with Fiberoptic and Digital Flexible Ureteroscopes at a High Volume Teaching Hospital

Date
2017-05
Language
English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Elsevier
Abstract

Introduction

Modern flexible ureteroscope ownership costs are considerable. Most prior estimates focus exclusively on repair costs, likely underestimating overall costs, including those of acquisition and reprocessing. Furthermore, to our knowledge no prior cost analyses focus on the latest generation digital flexible ureteroscope, which may differ due to unique purchase and repair prices. We sought to gain greater insight into the comprehensive costs associated with modern flexible ureteroscope use, particularly the difference between digital and fiberoptic models.

Methods

Data on use and repair of fiberoptic Storz Flex-X2 and digital Flex-Xc flexible ureteroscopes from 2011 to 2015 were reviewed. List prices and repair costs were obtained from Storz. Per case reprocessing costs were estimated, accounting for disposables, reagents and labor. Maintenance costs were estimated by combining cost of repairs and reprocessing. Analyses were performed at list pricing and standard discount rates. Global flexible ureteroscope costs were calculated to account for the cost of acquisition, repair and maintenance of a new scope during its first 100 uses.

Results

Global costs associated with digital flexible ureteroscope ownership were 1.3 to 1.4 times greater than fiberoptic on a per case basis ($1,008/$1,086 vs $715/$835). The majority of expenses went toward scope repairs (73% vs 71%), with instrument purchase (23% vs 24%) and reprocessing (4% vs 5%) being less costly. Repair rates were not significantly different between fiberoptic and digital devices (12.5 vs 11.5, p=0.757).

Conclusions

Expenditures associated with ownership of modern flexible ureteroscopes are considerable and driven primarily by the high cost of repairs. Digital instruments are more costly despite comparable rates of flexible ureteroscope damage.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Borofsky, M. S., Dauw, C. A., York, N. E., Hoovler, C., & Lingeman, J. E. (2017). Comprehensive Costs Associated with Fiberoptic and Digital Flexible Ureteroscopes at a High Volume Teaching Hospital. Urology Practice, 4(3), 187-192. http://dx.doi.org/10.1016/j.urpr.2016.06.007
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Urology Practice
Rights
Publisher Policy
Source
Author
Alternative Title
Type
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}