Failure Mode Analysis of the Endologix Endograft

Date
2016-09
Language
English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Elsevier
Abstract

Objective Type III (T-III) endoleaks following endovascular aneurysm repair (EVAR) remain a major concern. Our center experienced a recent concentration of T-III endoleaks requiring elective and emergency treatment and prompted our review of all EVAR implants over a 40-month period from April 2011 until August 2014. This report represents a single center experience with T-III endoleak management with analysis of factors leading to the T-III-related failure of EVAR.

Methods A retrospective review of all the operative reports, medical records, and computed tomography scans were reviewed from practice surveillance. Using Society for Vascular Surgery aneurysm reporting standards, we analyzed the morphology of the aneurysms before and after EVAR implant using computed tomography. Index procedure and frequency of reinterventions required to maintain aneurysm freedom from rupture were compared across all devices using SAS v 9.4 (SAS Institute, Inc, Cary, NC). Major adverse events (MAEs) requiring secondary interventions for aneurysm treatment beyond primary implant were analyzed for methods of failure. Aneurysm morphology of patients requiring EVAR was compared across all endograft devices used for repair. For purposes of MAE analysis, patients receiving Endologix (ELX) endograft were combined into group 1; Gore, Cook, and Medtronic endograft patients were placed into group 2.

Results Overall, technical success and discharge survival were achieved in 97.3% and 98% of patients regardless of device usage. There was no significant device related difference identified between patient survival or freedom from intervention. MAEs involving aneurysm treatment were over seven-fold more frequent with ELX (group 1) vs non-ELX (group 2) endografts (P < .01). Group 1 patients with aneurysm diameters larger than 65 mm were associated with a highly significant value for development of a T-III endoleak (odds ratio, 11.16; 95% confidence interval, 2.17, 57.27; P = .0038).

Conclusions While EVAR technical success and survival were similar across all devices, ELX devices exhibited an unusually high incidence of T-III endoleaks when implanted in abdominal aortic aneurysms with a diameter of more than 65 mm. Frequent reinterventions were required for Endologix devices for prevention of aneurysm rupture due to T-III endoleaks.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Lemmon, G. W., Motaganahalli, R. L., Chang, T., Slaven, J., Aumiller, B., Kim, B. J., & Dalsing, M. C. (2016). Failure mode analysis of the Endologix endograft. Journal of Vascular Surgery, 64(3), 571–576. https://doi.org/10.1016/j.jvs.2016.03.416
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Journal of Vascular Surgery
Rights
Publisher Policy
Source
Author
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}