Objective To determine if an innovative extracorporeal electrohydraulic shock wave device (sparker array) can effectively fracture artificial stones in vitro and in vivo, and if sparker array treatment produces a renal lesion in our pig model of lithotripsy injury. Results of these experiments will be used to help evaluate the suitability of this device as a clinical lithotripter.

Methods Utracal-30 artificial stones were placed in a holder at the focus of the sparker array and treated with 600 shock waves (21.6 kV, 60 shocks/min). Stone fragments were collected, dried and weighed to determine stone breakage. In vivo stone breakage entailed implanting stones into pigs. These stones were treated with 600 or 1200 shock waves and the fragments collected for analysis. Lesion analysis consisted of treating the left kidney of pigs with 1200 or 2400 shock waves and quantitating the hemorrhagic lesion.

Results In vitro, 71±2% of each artificial stone was fractured to < 2 mm in size. In vivo stone breakage averaged 63%. Renal injury analysis revealed that only 1 out of 7 kidneys showed evidence of hemorrhagic injury in the treated area.

Conclusions The sparker array consistently comminuted artificial stones demonstrating its ability to fracture stones like other lithotripters. Also, the sparker array caused little to no renal injury at the settings used in this study. These findings suggest further research is warranted to determine the potential of this device as a clinical lithotripter.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Connors, B. A., Schaefer, R. B., Gallagher, J. J., Johnson, C. D., Li, G., Handa, R. K., & Evan, A. P. (2018). Preliminary Report on Stone Breakage and Lesion Size Produced by a New Extracorporeal Electrohydraulic (Sparker Array) Discharge Device. Urology. https://doi.org/10.1016/j.urology.2018.03.020
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Urology
Rights
Publisher Policy
Source
Author
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}