Renal Protection Phenomenon Observed in a Porcine Model After Electromagnetic Lithotripsy Using a Treatment Pause
Date
Language
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Abstract
Purpose: Pretreating the kidney with 100 low-energy shock waves (SWs) with a time pause before delivering a clinical dose of SWs (Dornier HM-3, 2000 SWs, 24 kV, and 120 SWs/min) has been shown to significantly reduce the size of the hemorrhagic lesion produced in that treated kidney, compared to a protocol without pretreatment. It has been assumed that a similar reduction in injury will occur with lithotripters other than the HM-3, but experiments to confirm this assumption are lacking. In this study, we sought to verify that the lesion protection phenomenon also occurs in a lithotripter using an electromagnetic shock source and dry-head coupling.
Materials and Methods: Eleven female pigs were placed in a Dornier Compact S lithotripter where the left kidney of each animal was targeted for lithotripsy treatment. Some kidneys received 2500 SWs at power level (PL) = 5 (120 SWs/min) while some kidneys were pretreated with 100 SWs at PL = 1, with a 3-minute time pause, followed immediately by 2500 SWs at PL = 5 (120 SWs/min). Lesion size analysis was performed to assess the volume of hemorrhagic tissue injury created by each treatment regimen (% functional renal volume).
Results: Lesion size fell by 85% (p = 0.01) in the 100 SW pretreatment group compared to the injury produced by a regimen without pretreatment.
Conclusions: The results suggest that the treatment pause protection phenomenon occurs with a Dornier Compact S, a machine distinctly different from the Dornier HM-3. This result also suggests that this phenomenon may be observed generally in SW lithotripters.