Critical flaws in the T-Scan digital occlusion analysis system

dc.contributor.authorKatona, Thomas R.
dc.date.accessioned2022-09-09T18:39:36Z
dc.date.available2022-09-09T18:39:36Z
dc.date.issued2022-09-08
dc.description.abstractAbstract The T-Scan digital occlusion analysis system has previously been criticized for its use of artefactual measurements. Far more than the traditional clinical occlusion detecting armamentaria (paper, film and silk), its crown-sensor-crown configured force measurements misrepresent the actual crown-crown contact forces. Yet another defect, thus far unaddressed, is the misapplication of basic Newtonian static equilibrium principles, presumably the basis of their (proprietary) numerical algorithm. The approach requires the complete characterization (magnitudes, directions and points of application (or the lines-of-action)) of the occlusal contact force vectors. Instead, because of inherent sensor limitations, their elaborate mechanics calculations rely on artefactual force magnitudes, assumed directions, and estimated contact locations. Thus, the seemingly impressive T-Scan analyses are, essentially, groundless.en_US
dc.identifier.urihttps://hdl.handle.net/1805/29967
dc.language.isoen_USen_US
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 International*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectOcclusionen_US
dc.subjectT-Scanen_US
dc.titleCritical flaws in the T-Scan digital occlusion analysis systemen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
T-Scan 220908 A .pdf
Size:
195.75 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: