Superimposition of beams to vary shot size in gamma knife stereotactic radiosurgery

dc.contributor.authorBank, Morris I.
dc.contributor.authorTimmerman, Robert
dc.contributor.departmentRadiation Oncology, School of Medicineen_US
dc.date.accessioned2018-08-13T14:59:42Z
dc.date.available2018-08-13T14:59:42Z
dc.date.issued2002
dc.description.abstractThe Leksell Gamma Knife [Elekta Corp] uses helmets as collimators to produce four standard beam sizes. The nominal beam diameters are 18, 14, 8, and 4 mm. During computer treatment planning for gamma knife stereotactic radiosurgery, the size of the treated volume may differ from the standard beam sizes. To maintain conformality of the isodose curves to the treated volume, beam sizes may be superimposed during computer treatment planning to produce beam diameters that are intermediate to the standard beams. A study of superimposed gamma knife beams was performed to confirm the accuracy of this method and to verify the accuracy of the GammaPlan treatment planning computer. Superimposed beams were simulated on the Elekta treatment planning computer, GammaPlan, version 4.12, and tested by film measurements of beam profiles for single helmet sizes and superimposed shots with various beam weightings. The weighting for each beam size is varied to attain the beam size diameter desired. The beams were defined at the 50% isodose line. The profiles of the superimposed beams were obtained and compared with the single helmet shots. The uniformity of the resulting beams was measured. The results show a linear relationship between beam size and beam weighting for the superimposed beams. The film measurements confirm the computer calculations.en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.citationBank, M. I., & Timmerman, R. (2002). Superimposition of beams to vary shot size in gamma knife stereotactic radiosurgery. Journal of Applied Clinical Medical Physics, 3(1), 19–25. http://doi.org/10.1120/jacmp.v3i1.2588en_US
dc.identifier.urihttps://hdl.handle.net/1805/17113
dc.language.isoen_USen_US
dc.publisherAmerican Association of Physicists in Medicineen_US
dc.relation.isversionof10.1120/jacmp.v3i1.2588en_US
dc.relation.journalJournal of Applied Clinical Medical Physicsen_US
dc.rightsAttribution 3.0 United States
dc.rights.urihttp://creativecommons.org/licenses/by/3.0/us/
dc.sourcePMCen_US
dc.subjectGamma knifeen_US
dc.subjectRadiation therapyen_US
dc.subjectBeam profilesen_US
dc.titleSuperimposition of beams to vary shot size in gamma knife stereotactic radiosurgeryen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ACM2-3-19.pdf
Size:
94.63 KB
Format:
Adobe Portable Document Format
Description:
Main article
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: