Abstract

Objective To compare differences in audiologic outcomes between slim modiolar electrode (SME) CI532 and slim lateral wall electrode (SLW) CI522 cochlear implant recipients. Study Design Retrospective cohort study. Setting Tertiary academic hospital. Methods Comparison of postoperative AzBio sentence scores in quiet (percentage correct) in adult cochlear implant recipients with SME or SLW matched for preoperative AzBio sentence scores in quiet and aided and unaided pure tone average. Results Patients with SLW (n = 52) and patients with SME (n = 37) had a similar mean (SD) age (62.0 [18.2] vs 62.6 [14.6] years, respectively), mean preoperative aided pure tone average (55.9 [20.4] vs 58.1 [16.4] dB; P = .59), and mean AzBio score (percentage correct, 11.1% [13.3%] vs 8.0% [11.5%]; P = .25). At last follow-up (SLW vs SME, 9.0 [2.9] vs 9.9 [2.6] months), postoperative mean AzBio scores in quiet were not significantly different (percentage correct, 70.8% [21.3%] vs 65.6% [24.5%]; P = .29), and data log usage was similar (12.9 [4.0] vs 11.3 [4.1] hours; P = .07). In patients with preoperative AzBio <10% correct, the 6-month mean AzBio scores were significantly better with SLW than SME (percentage correct, 70.6% [22.9%] vs 53.9% [30.3%]; P = .02). The intraoperative tip rollover rate was 8% for SME and 0% for SLW. Conclusions Cochlear implantation with SLW and SME provides comparable improvement in audiologic functioning. SME does not exhibit superior speech recognition outcomes when compared with SLW.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
MacPhail, M. E., Connell, N. T., Totten, D. J., Gray, M. T., Pisoni, D., Yates, C. W., & Nelson, R. F. (2022). Speech Recognition Outcomes in Adults With Slim Straight and Slim Modiolar Cochlear Implant Electrode Arrays. Otolaryngology–Head and Neck Surgery, 166(5), 943–950. https://doi.org/10.1177/01945998211036339
ISSN
0194-5998, 1097-6817
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Otolaryngology–Head and Neck Surgery
Rights
Publisher Policy
Source
Author
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}