In-vivo application of low frequency alternating currents on porcine cervical vagus nerve evokes reversible nerve conduction block

dc.contributor.authorMuzquiz, Maria Ivette
dc.contributor.authorRichardson, Lindsay
dc.contributor.authorVetter, Christian
dc.contributor.authorSmolik, Macallister
dc.contributor.authorAlhawwash, Awadh
dc.contributor.authorGoodwill, Adam
dc.contributor.authorBashirullah, Rizwan
dc.contributor.authorCarr, Michael
dc.contributor.authorYoshida, Ken
dc.contributor.departmentBiomedical Engineering, School of Engineering and Technologyen_US
dc.date.accessioned2022-12-21T13:29:40Z
dc.date.available2022-12-21T13:29:40Z
dc.date.issued2021-06-30
dc.description.abstractBackground: This paper describes a method to reversibly block nerve conduction through direct application of a 1 Hz sinusoidal current waveform delivered through a bipolar nerve cuff electrode. This low frequency alternating current (LFAC) waveform was previously shown to reversibly block the effects of vagal pulse stimulation evoked bradycardia in-vivo in the anaesthetised rat model (Mintch et al. 2019). The present work measured the effectiveness of LFAC block on larger caliber myelinated vagal afferent fibers in human sized nerve bundles projecting to changes in breathing rate mediated by the Hering-Breuer (HB) reflex in anaesthetized domestic swine (n=5). Methods: Two bipolar cuff electrodes were implanted unilaterally to the left cervical vagus nerve, which was crushed caudal to the electrodes to eliminate cardiac effects. A tripolar recording cuff electrode was placed rostral to the bipolar stimulating electrodes on the same nerve to measure changes in the compound nerve action potentials (CNAP) elicited by the vagal pulse stimulation and conditioned by the LFAC waveform. Standard pulse stimulation was applied at a sufficient level to induce a reduction in breathing rate through the HB reflex. If unblocked, the HB reflex would cause breathing to slow down and potentially halt completely. Block was quantified by the ability of LFAC to reduce the effect of the HB reflex by monitoring the respiration rate during LFAC alone, LFAC and vagal stimulation, and vagal stimulation alone. Results: LFAC achieved 87.2 ±8.8% block (n=5) at current levels of 1.1 ±0.3 mAp (current to peak), which was well within the water window of the working electrode. CNAP showed changes that directly correlated to the effectiveness of LFAC block, which manifested itself as the slowing and amplitude reduction of components of the CNAP. Conclusion: These novel findings suggest that LFAC is a potential alternative or complementary method to other electrical blocking techniques in clinical applications.en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.citationMuzquiz MI, Richardson L, Vetter C, et al. In-vivo application of low frequency alternating currents on porcine cervical vagus nerve evokes reversible nerve conduction block. Bioelectron Med. 2021;7(1):9. Published 2021 Jun 30. doi:10.1186/s42234-021-00072-wen_US
dc.identifier.urihttps://hdl.handle.net/1805/30778
dc.language.isoen_USen_US
dc.publisherBMCen_US
dc.relation.isversionof10.1186/s42234-021-00072-wen_US
dc.relation.journalBioelectronic Medicineen_US
dc.rightsAttribution 4.0 International*
dc.rights.urihttps://creativecommons.org/licenses/by/4.0*
dc.sourcePMCen_US
dc.subjectNeuromodulationen_US
dc.subjectConduction blocken_US
dc.subjectNerve blocken_US
dc.subjectLow frequency alternating current blocken_US
dc.titleIn-vivo application of low frequency alternating currents on porcine cervical vagus nerve evokes reversible nerve conduction blocken_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
42234_2021_Article_72.pdf
Size:
2.97 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: