Simple, Inexpensive Attainment and Measurement of Very High Cooling and Warming Rates
Date
2010
Language
American English
Embargo Lift Date
Department
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Elsevier
Can't use the file because of accessibility barriers? Contact us with the title of the item, permanent link, and specifics of your accommodation need.
Abstract
We have developed a simple, inexpensive system (<$300 US) for measuring cooling and warming rates of small (∼ 0.1μl) aqueous samples at rates as high as 10(5)°C/min. The measurement system itself, can track rates approaching one million°C/min. For temperature sensing, a Type T thermocouple with 50μm wire was used. The thermocouple output voltage was read with an inexpensive USB based digital oscilloscope interfaced to a laptop computer, and the raw data were processed with MS Excel.
Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Kleinhans FW, Seki S, Mazur P. Simple, inexpensive attainment and measurement of very high cooling and warming rates. Cryobiology. 2010;61(2):231-233. doi:10.1016/j.cryobiol.2010.06.011