Cross-Polarization Optical Coherence Tomographic Assessment of In situ Simulated Erosive Tooth Wear
Date
Language
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Abstract
This clinical study tested cross-polarization optical coherence tomography (CP-OCT) monitoring of erosive tooth wear (ETW). Twenty participants completed a 14-day/arm, 3-arm crossover study simulating different ETW severities. Participants received two enamel specimens (per arm) and were randomized to: severe (s-ETW, lemon juice/pH:2.5/4.25%wt/vol citric acid), moderate (m-ETW, grapefruit juice/pH:3.5/1.03%wt/vol citric acid), and non-ETW (water). Enamel thickness was measured with CP-OCT (day[D] 0, 7, 14) and micro-computed tomography (μ-CT; D14). Enamel surface loss was determined with CP-OCT and optical profilometry (OP; D7, D14). CP-OCT showed higher enamel surface loss for D14 than D7 for m-ETW (P = .009) and s-ETW (P = .040) and differentiated severity at D14 (s-ETW > non-ETW, P = .027). OP was able to differentiate surface loss between days (D7 < D14, P < .001) for m-ETW and s-ETW, and ETW severity effect after 7 and 14 days (non-ETW < m-ETW < s-ETW, P < .001). At D14, CP-OCT and μ-CT were positively correlated (r = .87, ICC = .62). CP-OCT showed potential as a tool for clinical ETW monitoring.