Pediatric phantom dosimetry using a hand-held portable dental radiology device

dc.contributor.authorBozic, Zachary D.
dc.contributor.authorYepes, Juan F.
dc.contributor.authorJones, James E.
dc.contributor.authorSanders, Brian J.
dc.contributor.authorVinson, LaQuia
dc.contributor.authorJohnson, Brandon
dc.contributor.departmentPediatric Dentistry, School of Dentistryen_US
dc.date.accessioned2021-10-28T18:23:48Z
dc.date.available2021-10-28T18:23:48Z
dc.date.issued2020
dc.description.abstractBackground: The purpose of the study was to evaluate the radiation dose of the NOMAD Pro 2 device for different anatomical areas using a pediatric phantom. Methods: Absorbed doses resulting from a maxillary anterior occlusal and bitewing projections of an anthropomorphic 10-year-old child phantom were acquired using optical stimulated dosimetry. Equivalent doses were calculated for radiosensitive tissues in the head and neck area, and effective dose for maxillary anterior occlusal and bitewing examinations were calculated following the 2007 recommendations of the International Commission on Radiological Protection (ICRP). In addition, the effective dose of backscatter radiation to the operator was recorded. Results: Of the anterior occlusal scans, the salivary glands had the highest equivalent dose, followed by oral mucosa extra thoracic airway and thyroid gland. For the bitewing projection scan, the salivary glands had the highest equivalent dose followed closely by the oral mucosa. The operator had minimal recorded dose. Conclusion: Compared to previous research, completed with the adult phantom, a child receives less radiation for bitewing and anterior occlusal projections using a hand-held portable device than when a traditional wall mounted device is utilized.en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.citationBozic, Z., Yepes, J., Jones, J., Sanders, B., Vinson, L., & Johnson, B. (2020). Pediatric phantom dosimetry using a hand-held portable dental radiology device. Annals of Dentistry and Oral Health, 3(1), 1012.en_US
dc.identifier.urihttps://hdl.handle.net/1805/26891
dc.language.isoenen_US
dc.publisherMedDocsen_US
dc.relation.journalAnnals of Dentistry and Oral Healthen_US
dc.rightsAttribution 4.0 International*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/*
dc.sourcePublisheren_US
dc.subjectradiation doseen_US
dc.subjectphantom dosimetryen_US
dc.subjecthand-held portable dental radiology deviceen_US
dc.titlePediatric phantom dosimetry using a hand-held portable dental radiology deviceen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Bozic2020Pediatric-CCBY.pdf
Size:
632.63 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: