Dysfunctional Implantable Pulmonary Artery Sensor Device (CardioMEMS) in Group 2 Pulmonary Hypertension

dc.contributor.authorAbdelkader, Abdalla Eltayeb A.
dc.contributor.authorIlonze, Onyedika J.
dc.contributor.authorGuglin, Maya
dc.contributor.departmentMedicine, School of Medicine
dc.date.accessioned2025-02-17T16:34:28Z
dc.date.available2025-02-17T16:34:28Z
dc.date.issued2025-01-15
dc.description.abstractImplantable hemodynamic devices like the CardioMEMS HF System are commonly used to manage fluid status in patients with heart failure (HF) by measuring pulmonary pressures. Although CardioMEMS has been shown to reduce HF hospitalizations, rare complications (eg, device endothelialization) can occur and warrant clinical attention. A 67-year-old woman with HF with preserved ejection fraction and group 2 pulmonary hypertension experienced recurrent HF exacerbations. Despite optimal therapy, she was not a candidate for advanced HF therapies. The CardioMEMS device, initially effective for fluid management, showed dampened waveforms due to endothelialization, leading to reimplantation. Endothelialization is a rare but significant complication that can dampen pressure waveforms. Proper placement in vessels larger than 7 mm and careful monitoring of waveforms can help manage this issue. Device recalibration can usually address most cases; however, reimplantation may be required.
dc.eprint.versionFinal published version
dc.identifier.citationAbdelkader AEA, Ilonze OJ, Guglin M. Dysfunctional Implantable Pulmonary Artery Sensor Device (CardioMEMS) in Group 2 Pulmonary Hypertension. JACC Case Rep. 2025;30(2):102801. Published 2025 Jan 15. doi:10.1016/j.jaccas.2024.102801
dc.identifier.urihttps://hdl.handle.net/1805/45744
dc.language.isoen_US
dc.publisherElsevier
dc.relation.isversionof10.1016/j.jaccas.2024.102801
dc.relation.journalJACC Case Reports
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internationalen
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0
dc.sourcePMC
dc.subjectCardioMEMS
dc.subjectDampened waveform
dc.subjectHeart failure with preserved ejection fraction
dc.subjectPulmonary hypertension
dc.titleDysfunctional Implantable Pulmonary Artery Sensor Device (CardioMEMS) in Group 2 Pulmonary Hypertension
dc.typeArticle
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Abdelkader2025Dysfunctional-CCBYNCND.pdf
Size:
1.12 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2.04 KB
Format:
Item-specific license agreed upon to submission
Description: