Rapid quantification of underivatized alloisoleucine and argininosuccinate using mixed-mode chromatography with tandem mass spectrometry

Date
2019-10
Language
English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Elsevier
Abstract

Plasma elevations of the amino acids alloisoleucine and argininosuccinic acid (ASA) are pathognomonic for maple syrup urine disease and argininosuccinate lyase deficiency, respectively. Reliable detection of these biomarkers is typically achieved using methods with tedious sample preparations or long chromatographic separations, and many published amino acid assays report poor specificity and/or sensitivity for one or both of these compounds. This report describes a novel liquid chromatography tandem mass spectrometry (LC-MS/MS) method that provides rapid quantification of alloisoleucine and ASA in human plasma. The basis of this method is a mixed-mode solid phase separation that achieves baseline resolution of alloisoleucine from isobaric interferents without the use of derivatization or ion pairing agents. The inject-to-inject time is 6 min including elution, column washing and re-equilibration. Validation studies demonstrate excellent limits of quantification (1 μmol/L), linearity (r = 0.999 from 1 to 250 μmol/L), accuracy (bias = −3.8% and −10.1%), and inter-assay imprecision (CV < 8.06%) for plasma analyses. Data from long-term clinical application confirms chromatographic consistency equivalent to more traditional reversed-phase or HILIC based columns. Additional matrix studies indicate low suppression (<10%) for a wide range of amino acids and compatibility with other matrixes such as blood spot analyses. Finally, analysis of our first 257 clinical specimens demonstrates high analytic specificity and sensitivity, allowing the detection of subtle but clinically relevant elevations of alloisoleucine and ASA that may be missed by other less sensitive methods. In conclusion, the novel LC-MS/MS method reported here overcomes a number of the challenges associated with alloisoleucine and ASA quantification. Combining this approach with published incomplete amino acid quantification methods allows, for the first time, a rapid and comprehensive LC-MS/MS analysis of underivatized amino acids without the use of ion pairing agents.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Griffin, C., Ammous, Z., Vance, G. H., Graham, B. H., & Miller, M. J. (2019). Rapid quantification of underivatized alloisoleucine and argininosuccinate using mixed-mode chromatography with tandem mass spectrometry. Journal of Chromatography B, 1128, 121786. https://doi.org/10.1016/j.jchromb.2019.121786
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Journal of Chromatography B
Rights
Publisher Policy
Source
Publisher
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}