Computational simulations establish a novel transducer array placement arrangement that extends delivery of therapeutic TTFields to the infratentorium of patients with brainstem gliomas

If you need an accessible version of this item, please email your request to digschol@iu.edu so that they may create one and provide it to you.
Date
2021-12-30
Language
American English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Elsevier
Abstract

Background and purpose: Tumor treating fields (TTFields) are a non-invasive, efficacious treatment modality currently approved for supratentorial glioblastomas. Despite their ability to improve overall survival in supratentorial tumors, the current placement of arrays is limited to the supratentorial head, precluding its use in infratentorial tumors. Infratentorial malignancies are in need of new therapy modalities given their poor prognoses in both children and adults. The aim of this research is to determine whether rearrangement of TTFields may allow for management of infratentorial tumors.

Materials and methods: Delivery of TTFields using Novocure's prototype Optune™ device human male head model was simulated based on brain MRIs from patients with brainstem gliomas to develop a novel array layout designed to extend adequate infratentorial coverage.

Results: Array placement on the vertex, bilateral posterolateral occiput, and superior-posterior neck achieved intensities above 1.1 V/cm (average 1.7 V/cm; maximum 2.3 V/cm) in the vertical field direction and above 1 V/cm (average 2 V/cm; maximum 2.8 V/cm) in the horizontal field direction of the infratentorium. The calculated field intensity within the simulated tumors were in the therapeutic range and demonstrated the effective delivery of TTFields to the infratentorial brain.

Conclusions: Our findings suggest that rearrangement of the TTFields standard array with placement of electrodes on the vertex, bilateral posterolateral occiput, and superior-posterior neck allows for adequate electric field distribution in the infratentorium that is within the therapeutic range.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Ramirez-Fort MK, Naveh A, McClelland S 3rd, et al. Computational simulations establish a novel transducer array placement arrangement that extends delivery of therapeutic TTFields to the infratentorium of patients with brainstem gliomas. Rep Pract Oncol Radiother. 2021;26(6):1045-1050. Published 2021 Dec 30. doi:10.5603/RPOR.a2021.0132
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Reports of Practical Oncology and Radiotherapy
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Final published version
Full Text Available at
This item is under embargo {{howLong}}