Mechatronic Design and Active Disturbance Rejection Control of a Bag Valve-Based Mechanical Ventilator

dc.contributor.authorArcos-Legarda, Jaime
dc.contributor.authorTovar, Andres
dc.contributor.departmentMechanical Engineering, School of Engineering and Technologyen_US
dc.date.accessioned2023-06-16T21:00:37Z
dc.date.available2023-06-16T21:00:37Z
dc.date.issued2021-09
dc.description.abstractThis paper presents the mechatronic (mechanical and control system) design of a functional prototype of a portable mechanical ventilator to treat patients with a compromised respiratory function. The portable ventilator ensures adequate oxygenation and carbon dioxide clearance while avoiding ventilator-induced lung injury (VILI). Oxygen is delivered through the compression of a bag valve (Ambu bag) using a moving strap. Carbon dioxide is cleared through the control of a pinch valve actuated by a low-torque servomotor. The positive end-expiratory pressure (PEEP) is controlled by an adjustable mechanical valve of the system. An Arduino Mega microcontroller board is used in this prototype to control the respiratory variables. All mechanical components as well as sensors, actuators, and control hardware are of common use in robotics and are very inexpensive. The total cost of the prototype built in this work is about $425 U.S. dollars. The design is meant to be replicated and utilized in emergency conditions that involve an overwhelming number of cases, such as COVID-19 treatment, in places with no access to commercial mechanical ventilation (MV) technologies. In order to account for variations in the prototype as built, the software developed for this portable MV applies an active disturbance rejection control (ADRC) strategy. This control strategy is presented as a universal control structure for any mechanical ventilator able to supply air flow with controlled pressure and volume.en_US
dc.eprint.versionAuthor's manuscripten_US
dc.identifier.citationArcos-Legarda, J., & Tovar, A. (2021). Mechatronic Design and Active Disturbance Rejection Control of a Bag Valve-Based Mechanical Ventilator. Journal of Medical Devices, 15(3), 031006. https://doi.org/10.1115/1.4051064en_US
dc.identifier.issn1932-6181, 1932-619Xen_US
dc.identifier.urihttps://hdl.handle.net/1805/33840
dc.language.isoen_USen_US
dc.publisherASMEen_US
dc.relation.isversionof10.1115/1.4051064en_US
dc.relation.journalJournal of Medical Devicesen_US
dc.rightsPublisher Policyen_US
dc.sourceAuthoren_US
dc.subjectEngineering prototypesen_US
dc.subjectCompression Hardwareen_US
dc.subjectDesignen_US
dc.subjectVentilationen_US
dc.titleMechatronic Design and Active Disturbance Rejection Control of a Bag Valve-Based Mechanical Ventilatoren_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Arcos-Legarda2021Mechatronic-AAM.pdf
Size:
1.32 MB
Format:
Adobe Portable Document Format
Description:
Article
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: