Oxygenation Profiles of Human Blood, Cell Culture Medium, and Water for Perfusion of 3D-Bioprinted Tissues using the FABRICA Bioreactor Platform

If you need an accessible version of this item, please email your request to digschol@iu.edu so that they may create one and provide it to you.
Date
2020-04-29
Language
English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Nature
Abstract

Persistent and saturated oxygen distribution from perfusion media (i.e., blood, or cell culture media) to cells within cell-dense, metabolically-active biofabricated tissues is required to keep them viable. Improper or poor oxygen supply to cells within the tissue bulk severely limits the tissue culturing potential of many bioreactors. We added an oxygenator module to our modular FABRICA bioreactor in order to provide stable oxygenation to biofabricated tissues during culture. In this proof of concept study of an oxygenated and perfused bioreactor, we characterized the oxygenation of water, cell culture medium, and human blood in the FABRICA as functions of augmenting vacuum (air inlet) pressure, perfusion (volumetric flow) rate, and tubing/oxygenator components. The mean oxygen levels for water and cell culture media were 27.7 ± 2.1% and 27.6 ± 4.1%, respectively. The mean oxygen level for human blood was 197.0 ± 90.0 mmHg, with near-physiologic levels achieved with low-permeability PharMed tubing alone (128.0 ± 14.0 mmHg). Hematologic values pre- and post-oxygenation, respectively were (median ± IQR): Red blood cell: 6.0 ± 0.5 (106/μL) and 6.5 ± 0.4 (106/μL); Hemoglobin: 17.5 ± 1.2 g/dL and 19.2 ± 3.0 g/dL; and Hematocrit: 56.7 ± 2.4% and 61.4 ± 7.5%. The relative stability of the hematologic parameters indicates that blood function and thus blood cell integrity were maintained throughout oxygenation. Already a versatile research tool, the now oxygenated FABRICA provides easy-to-implement, in vivo-like perfusion and stable oxygenation culture conditions in vitro semi-independently of one another, which means the bioreactor has the potential to serve as a platform for investigating the behavior of 3D tissue models (regardless of biofabrication method), performing drug toxicity-testing, and testing pharmaceutical efficacy/safety.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Chen, A. M., Lashmet, M., Isidan, A., Sterner, J. L., Walsh, J., Koehler, C., Li, P., Ekser, B., & Smith, L. (2020). Oxygenation Profiles of Human Blood, Cell Culture Medium, and Water for Perfusion of 3D-Bioprinted Tissues using the FABRICA Bioreactor Platform. Scientific Reports, 10(1), 7237. https://doi.org/10.1038/s41598-020-64256-1
ISSN
2045-2322
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Scientific Reports
Source
Publisher
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Final published version
Full Text Available at
This item is under embargo {{howLong}}