Performance of a 62Zn/62Cu microgenerator in kit-based synthesis and delivery of [62Cu]Cu–ETS for PET perfusion imaging

dc.contributor.authorNg, Yen
dc.contributor.authorLacy, Jeffrey L.
dc.contributor.authorFletcher, James W.
dc.contributor.authorGreen, Mark A.
dc.contributor.departmentDepartment of Radiology and Imaging Sciences, IU School of Medicineen_US
dc.date.accessioned2016-02-29T22:19:11Z
dc.date.available2016-02-29T22:19:11Z
dc.date.issued2014-09
dc.description.abstractThe performance of a commercially produced 62Zn/62Cu microgenerator system, and an associated kit-based radiopharmaceutical synthesis method, was evaluated for clinical site production of [62Cu]Cu-ETS (ethylglyoxal bis(thiosemicarbazonato)copper(II)), an investigational agent for PET perfusion imaging. Using 37 generators, containing 1.84 ± 0.23 GBq 62Zn at 9:00 AM on the day of clinical use, a total of 45 patient doses of [62Cu]Cu-ETS (672 ± 172 MBq) were delivered without difficulty. 62Cu elution yields were high (approximately 90%), accompanied by extremely low 62Zn breakthrough (<0.001%). Radiopharmaceutical preparation, from the start-of-elution to time-of-injection, consumed less than five minutes. The 62Zn/62Cu microgenerator was a dependable source of short-lived positron-emitting 62Cu, and the kit-based synthesis proved to be rapid, robust, and highly reliable for “on-demand” delivery of [62Cu]Cu-ETS for PET perfusion imaging.en_US
dc.eprint.versionAuthor's manuscripten_US
dc.identifier.citationNg, Y., Lacy, J. L., Fletcher, J. W., & Green, M. A. (2014). Performance of a 62Zn/62Cu Microgenerator in Kit-Based Synthesis and Delivery of [62Cu]Cu-ETS for PET Perfusion Imaging. Applied Radiation and Isotopes : Including Data, Instrumentation and Methods for Use in Agriculture, Industry and Medicine, 91, 38–43. http://doi.org/10.1016/j.apradiso.2014.05.006en_US
dc.identifier.urihttps://hdl.handle.net/1805/8594
dc.language.isoen_USen_US
dc.publisherElsevier B.V.en_US
dc.relation.isversionof10.1016/j.apradiso.2014.05.006en_US
dc.relation.journalApplied Radiation and Isotopesen_US
dc.rightsPublisher Policyen_US
dc.sourcePMCen_US
dc.subject62Zn/62Cu generatoren_US
dc.subjectPET radiopharmaceuticalsen_US
dc.subject[62Cu]Cu-ETS (copper(II) ethylglyoxal bis(thiosemicarbazone))en_US
dc.subjecttumor perfusionen_US
dc.titlePerformance of a 62Zn/62Cu microgenerator in kit-based synthesis and delivery of [62Cu]Cu–ETS for PET perfusion imagingen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
nihms599171.pdf
Size:
1.48 MB
Format:
Adobe Portable Document Format
Description:
Article
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.88 KB
Format:
Item-specific license agreed upon to submission
Description: