The SMART Registry: Long-Term Results on the Utility of the Penumbra SMART COIL System for Treatment of Intracranial Aneurysms and Other Malformations

Abstract

Introduction: Penumbra SMART COIL® (SMART) System is a novel generation embolic coil with varying stiffness. The study purpose was to report real-world usage of the SMART System in patients with intracranial aneurysms (ICA) and non-aneurysm vascular lesions.

Materials and Methods: The SMART Registry is a post-market, prospective, multicenter registry requiring ≥75% Penumbra Coils, including SMART, PC400, and/or POD coils. The primary efficacy endpoint was retreatment rate at 1-year and the primary safety endpoint was the procedural device-related serious adverse event rate.

Results: Between June 2016 and August 2018, 995 patients (mean age 59.6 years, 72.1% female) were enrolled at 68 sites in the U.S. and Canada. Target lesions were intracranial aneurysms in 91.0% of patients; 63.5% were wide-neck and 31.8% were ruptured. Adjunctive devices were used in 55.2% of patients. Mean packing density was 32.3%. Procedural device-related serious adverse events occurred in 2.6% of patients. The rate of immediate post-procedure adequate occlusion was 97.1% in aneurysms and the rate of complete occlusion was 85.2% in non-aneurysms. At 1-year, the retreatment rate was 6.8%, Raymond Roy Occlusion Classification (RROC) I or II was 90.0% for aneurysms, and Modified Rankin Scale (mRS) 0-2 was achieved in 83.1% of all patients. Predictors of 1-year for RROC III or retreatment (incomplete occlusion) were rupture status (P < 0.0001), balloon-assisted coiling (P = 0.0354), aneurysm size (P = 0.0071), and RROC III immediate post-procedure (P = 0.0086) in a model that also included bifurcation aneurysm (P = 0.7788). Predictors of aneurysm retreatment at 1-year was rupture status (P < 0.0001).

Conclusions: Lesions treated with SMART System coils achieved low long-term retreatment rates.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Spiotta AM, Park MS, Bellon RJ, et al. The SMART Registry: Long-Term Results on the Utility of the Penumbra SMART COIL System for Treatment of Intracranial Aneurysms and Other Malformations. Front Neurol. 2021;12:637551. Published 2021 Apr 13. doi:10.3389/fneur.2021.637551
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Frontiers in Neurology
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Final published version
Full Text Available at
This item is under embargo {{howLong}}