The effect of intraoral suction on oxygen-enriched surgical environments: a mechanism for reducing the risk of surgical fires
Date
Language
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Abstract
In this study, a mechanical model was applied in order to replicate potential surgical fire conditions in an oxygen-enriched environment with and without high-volume suction typical for dental surgical applications. During 41 trials, 3 combustion events were measured: an audible pop, a visible flash of light, and full ignition. In at least 11 of 21 trials without suction, all 3 conditions were observed, sometimes with an extent of fire that required early termination of the experimental trial. By contrast, in 18 of 20 with-suction trials, ignition did not occur at all, and in the 2 cases where ignition did occur, the fire was qualitatively a much smaller, candle-like flame. Statistically comparing these 3 combustion events in the no-suction versus with-suction trials, ignition (P = .0005), audible pop (P = .0211), and flash (P = .0092) were all significantly more likely in the no-suction condition. These results suggest a possible significant and new element to be added to existing surgical fire safety protocols toward making surgical fires the "never-events" they should be.