Evaluation of a Hybrid Iterative Reconstruction technique "iDose" for CT Dose Reduction and Image Quality

dc.contributor.authorLiang, Yun
dc.date.accessioned2014-10-22T15:56:06Z
dc.date.available2014-10-22T15:56:06Z
dc.date.issued2014-04-11
dc.descriptionposter abstracten_US
dc.description.abstractPurpose: To investigate the use of a commercial designed hybrid iterative reconstruction technique iDose (by Philips Medical Systems) for CT radiation dose reduction and its effect on image quality. Methods: A catphan®504 phantom was scanned using a 64-slice CT (Ingenuity, Philips Medical Systems) with two different tube voltages (100 kVp, and 120 kVp) at three different doses, a reference dose (CTDIvol) of 42 mGy and two reduced doses of about 50% and 75% of the reference. Images were reconstructed with standard filtered-back-projection (FBP) and with iDose algorithms. Six different iDose levels were employed. Quantitative evaluation of spatial resolution, image noise, noise power spectrum (NPS), and low-contrast detectability were carried out. Results: For any given dose level, there was a static noise reduction with increased iDose level over the FBP. To match the “standard” noise resulted from the FBP at the “reference dose”, a minimum iDose level of 4 and 6 was required for 50% and 75% dose reduction respectively. NPS showed moderate shift towards the lower frequency as the iDose level increased. The NPS shift was consistent with the observed subtle change of noise texture. This shift also correlated with the change in low-contrast detectability among images with the same noise level, the higher the iDose level that was used in image reconstruction, the lower the low-contrast detectability. Conclusions: The iDose algorithm clearly demonstrates effectiveness in noise suppression over the FBP. The low-contrast detectability depends on noise but also on NPS, which is shifted by iDose algorithm. At very low-dose levels, greater iDose levels would be needed to reduce the image noise but may not improve the low-contrast detectability. This findings indicate that for any given specific clinical task, the lowest dose limit achievable is determined by an “optimal” iDose level ascertaining the noise reduction as well as maintaining desired low-contrast detectability.en_US
dc.identifier.citationLiang, Y. (2014, April 11). Evaluation of a Hybrid Iterative Reconstruction technique "iDose" for CT Dose Reduction and Image Quality. Poster session presented at IUPUI Research Day 2014, Indianapolis, Indiana.en_US
dc.identifier.urihttps://hdl.handle.net/1805/5379
dc.language.isoen_USen_US
dc.publisherOffice of the Vice Chancellor for Researchen_US
dc.subjectiDoseen_US
dc.subjectCT radiation dose reductionen_US
dc.subjectimage qualityen_US
dc.titleEvaluation of a Hybrid Iterative Reconstruction technique "iDose" for CT Dose Reduction and Image Qualityen_US
dc.typeOtheren_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Liang-Evaluation.pdf
Size:
64.28 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.88 KB
Format:
Item-specific license agreed upon to submission
Description: