Pediatric Phantom Dosimetry of Kodak 9000 Cone-beam Computed Tomography

dc.contributor.authorYepes, Juan F.
dc.contributor.authorBooe, Megan R.
dc.contributor.authorSanders, Brian J.
dc.contributor.authorJones, James E.
dc.contributor.authorEhrlich, Ygal
dc.contributor.authorLudlow, John B.
dc.contributor.authorJohnson, Brandon
dc.contributor.departmentPediatric Dentistry, School of Dentistryen_US
dc.date.accessioned2018-03-07T17:43:53Z
dc.date.available2018-03-07T17:43:53Z
dc.date.issued2017-05
dc.description.abstractPurpose: The purpose of the study was to evaluate the radiation dose of the Kodak 9000 cone-beam computed tomography (CBCT) device for different anatomical areas using a pediatric phantom. Methods: Absorbed doses resulting from maxillary and mandibular region three by five cm CBCT volumes of an anthropomorphic 10-year-old child phantom were acquired using optical stimulated dosimetry. Equivalent doses were calculated for radiosensitive tissues in the head and neck area, and effective dose for maxillary and mandibular examinations were calculated following the 2007 recommendations of the International Commission on Radiological Protection (ICRP). Results: Of the mandibular scans, the salivary glands had the highest equivalent dose (1,598 microsieverts [μSv]), followed by oral mucosa (1,263 μSv), extrathoracic airway (pharynx, larynx, and trachea; 859 μSv), and thyroid gland (578 μSv). For the maxilla, the salivary glands had the highest equivalent dose (1,847 μSv), followed closely by oral mucosa (1,673 μSv), followed by the extrathoracic airway (pharynx, larynx, and trachea; 1,011 μSv) and lens of the eye (202 μSv). Conclusion: Compared to previous research of the Kodak 9000, completed with the adult phantom, a child receives one to three times more radiation for mandibular scans and two to 10 times more radiation for maxillary scans.en_US
dc.eprint.versionAuthor's manuscripten_US
dc.identifier.citationYepes, Juan F., Megan R. Booe, Brian J. Sanders, James E. Jones, Ygal Ehrlich, John B. Ludlow, and Brandon Johnson. "Pediatric Phantom Dosimetry of Kodak 9000 Cone-beam Computed Tomography." Pediatric dentistry 39, no. 3 (2017): 229-232.en_US
dc.identifier.urihttps://hdl.handle.net/1805/15399
dc.language.isoenen_US
dc.publisherAmerican Academy of Pediatric Dentistryen_US
dc.relation.journalPediatric Dentistryen_US
dc.rightsPublisher Policyen_US
dc.sourceAuthoren_US
dc.subjectchidrenen_US
dc.subjectcone beam computed tomographyen_US
dc.subjectdosimetryen_US
dc.titlePediatric Phantom Dosimetry of Kodak 9000 Cone-beam Computed Tomographyen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Yepes_2018_pediatric.pdf
Size:
268.34 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: