A Survey of Morcellator Preference and Cost Comparison of the Lumenis VersaCut and Wolf Piranha Morcellators

Date
2017
Language
English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Elsevier
Abstract

Objective

To evaluate operating room (OR) costs associated with the 2 available morcellators in the United States in a matched cohort and to determine benign prostatic hyperplasia surgeon's morcellator preference.

Materials and Methods

Patients from 2013, the last year our institution exclusively used the VersaCut device, were matched 1:1 with the most recent patient cohort, utilizing the Wolf Piranha morcellator. Cost of morcellation including the expense of OR time and disposable instrument costs were calculated. A survey to the Endourological Society e-mail listserv was sent to determine morcellator preference.

Results

We identified 142 patients who underwent holmium laser enucleation of the prostate in 2013. When compared with the VersaCut group, morcellation efficiency (4.4 vs 7.0 g/min, P <.01) and expense of OR time ($1420.80 vs $992.21, P <.005) both favored the Piranha morcellator system even when the costs of disposable instruments were factored into the analysis ($1338.81 vs $1637.50, P <.05).

A total of 126 urologists responded to the survey. Of these, 56 (44.5%) perform transurethral prostate enucleations, which included 48 (86%) holmium. More endourologists use the VersaCut (n = 33, 59%) than the Piranha (n = 24, 43%) morcellator. Qualities that impacted the preference of morcellator included the preferred device is safer, faster, easier to use, reusable, and less expensive.

Conclusion

We identified a significant improved efficiency and improved cost savings utilizing the Piranha morcellator even when controlling for disposable costs. Of the endourologists who responded to the survey, less than half perform transurethral enucleation. Morcellator preference is largely based on safety, efficiency, and ease of use, whereas cost and reusablility were of lesser importance.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Rivera, M. E., Lingeman, J. E., Heinsimer, K., York, N. E., & Krambeck, A. E. (2017). A Survey of Morcellator Preference and Cost Comparison of the Lumenis® VersaCut™ and Wolf Piranha Morcellators. Urology. https://doi.org/10.1016/j.urology.2017.09.019
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Urology
Rights
Publisher Policy
Source
Author
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}