A double-reprocessing high-level disinfection protocol does not eliminate positive cultures from the elevators of duodenoscopes

Date
2018
Language
English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Thieme
Abstract

Background and study aim Duodenoscopes have been the source of serious infection, despite correct performance of high-level disinfection (HLD). This study aimed to observe the impact of performing HLD twice on the rate of positive cultures from duodenoscope elevators.

Methods We performed double HLD (DHLD; i. e. complete manual cleaning followed by automated reprocessing, with the entire process repeated) and then randomly cultured the elevators of our duodenoscopes on about 30 % of occasions.

Results DHLD was associated with positive elevator cultures for any microorganism in 9.4 % of cases, with a 0.8 % rate of known pathogens (627 cultures) between May 2015 and February 2016. After February 2016, and in association with changing the precleaning fluid, as well as use of a new FDA-recommended cleaning brush, the rate of positive cultures for any microorganism after DHLD was 4.8 % and 0.2 % for known pathogens (420 cultures). In a third phase, characterized by a change in personnel performing DHLD and retirement of a duodenoscope with a high rate of positive cultures, the rate of positive cultures for any microorganism was 4.9 % (783 cultures) and the rate of positive culture for known pathogens was 0.3 %. To our knowledge, no duodenoscope transmission of infection occurred during the study interval.

Conclusions DHLD resulted in a low rate of positive cultures for known pathogens and for organisms of low pathogenic potential, but did not eliminate these, from duodenoscope elevators. Additional improvements in HLD protocols and/or duodenoscope design are needed.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Rex, D. K., Sieber, M., Lehman, G. A., Webb, D., Schmitt, B., Kressel, A. B., … Sherman, S. (2018). A double-reprocessing high-level disinfection protocol does not eliminate positive cultures from the elevators of duodenoscopes. Endoscopy, 50(6), 588–596. https://doi.org/10.1055/s-0043-122378
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Endoscopy
Rights
Publisher Policy
Source
Publisher
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Final published version
Full Text Available at
This item is under embargo {{howLong}}