Dose correction in lung for HDR breast brachytherapy

Date
2012
Language
American English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Termedia
Can't use the file because of accessibility barriers? Contact us with the title of the item, permanent link, and specifics of your accommodation need.
Abstract

Purpose: To evaluate the dosimetric impact of lung tissue in Ir-192 APBI.

Material and methods: In a 40 × 40 × 40 cm(3) water tank, an Accelerated Partial Breast Irradiation (APBI) brachytherapy balloon inflated to 4 cm diameter was situated directly below the center of a 30 × 30 × 1 cm(3) solid water slab. Nine cm of solid water was stacked above the 1 cm base. A parallel plate ion chamber was centered above the base and ionization current measurements were taken from the central HDR source dwell position for channels 1, 2, 3 and 5 of the balloon. Additional ionization data was acquired in the 9 cm stack at 1 cm increments. A comparable data set was also measured after replacing the 9 cm solid water stack with cork slabs. The ratios of measurements in the two phantoms were calculated and compared to predicted results of a commercial treatment planning system.

Results: Lower dose was measured in the cork within 1 cm of the cork/solid water interface possibly due to backscatter effects. Higher dose was measured beyond 1 cm from the cork/solid water interface, increasing with path length up to 15% at 9 cm depth in cork. The treatment planning system did not predict either dose effect.

Conclusions: This study investigates the dosimetry of low density material when the breast is treated with Ir-192 brachytherapy. HDR dose from Ir-192 in a cork media is shown to be significantly different than in unit density media. These dose differences are not predicted in most commercial brachytherapy planning systems. Empirical models based on measurements could be used to estimate lung dose associated with HDR breast brachytherapy.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Slessinger E, Pepin E, Zhao Q, Zhao L, Das I. Dose correction in lung for HDR breast brachytherapy. J Contemp Brachytherapy. 2012;4(2):106-110. doi:10.5114/jcb.2012.29367
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Journal of Contemporary Brachytherapy
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Final published version
Full Text Available at
This item is under embargo {{howLong}}