Pediatric Phantom Dosimetry Evaluation of a New Rectangular Collimator

dc.contributor.authorMartinez, Amanda
dc.contributor.authorYepes, Juan F.
dc.contributor.authorJones, James E.
dc.contributor.authorWong, Phillip
dc.contributor.authorJohnson, K. Brandon
dc.contributor.authorCanady, Sarah
dc.contributor.authorTang, Qing
dc.contributor.departmentPediatric Dentistry, School of Dentistry
dc.date.accessioned2025-04-04T20:11:22Z
dc.date.available2025-04-04T20:11:22Z
dc.date.issued2023
dc.description.abstractPurpose: To compare the effective dose (E) of the Tru-Image® rectangular collimator and the universal round collimator of a Planmeca® wall-mounted radiograph unit for two bitewing radiographs (right and left) on a pediatric phantom. Methods: Absorbed doses utilizing the Tru-Image ®rectangular collimator and universal round collimator were acquired using an anthropomorphic 10-year-old child phantom. Each set of 24 dosimeters was exposed to two bitewing exposures with the manufacturer's child settings. Fifty clinical exposures were completed for each set and three sets were exposed for each collimator. The average E per exposure was calculated. Results: The overall E for the Tru-Image ®rectangular collimator and the universal round collimator were 6.3 microsieverts (μSv) and 25.3 μSv, respectively. This difference was statistically significant (P <0.001). The highest equivalent dose for both collimators was delivered to the oral mucosa. When compared to the universal round collimator, the Tru-Image ® rectangular collimator had significant dose reduction at all locations (P <0.05). When normalized and adjusted to the same source-to-end distance, there was an overall 65 percent dose reduction with the rectangular collimator. Conclusion: The average effective dose was significantly reduced with the use of the Tru-Image ®rectangular collimator. Clinical use of this rectangular collimator should be considered in the pediatric population.
dc.eprint.versionAuthor's manuscript
dc.identifier.citationMartinez, A., Yepes, J. F., Jones, J. E., Wong, P., Johnson, K. B., Canady, S., & Tang, Q. (2023). Pediatric Phantom Dosimetry Evaluation of a New Rectangular Collimator. Journal of Dentistry for Children (Chicago, Ill.), 90(1), 3–10.
dc.identifier.urihttps://hdl.handle.net/1805/46847
dc.language.isoen
dc.publisherAmerican Academy of Pediatric Dentistry
dc.relation.journalJournal of Dentistry for Children
dc.rightsPublisher Policy
dc.sourceAuthor
dc.subjectdosimetry
dc.subjectrectangular collimator
dc.subjectbitewing
dc.titlePediatric Phantom Dosimetry Evaluation of a New Rectangular Collimator
dc.typeArticle
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Martinez2023Pediatric-AAM.pdf
Size:
515.54 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2.04 KB
Format:
Item-specific license agreed upon to submission
Description: