GRASP-Pro: imProving GRASP DCE‐MRI through self-calibrating subspace-modeling and contrast phase automation

Date
2020-01
Language
American English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Wiley
Abstract

Purpose: To propose a highly accelerated, high-resolution dynamic contrast-enhanced MRI (DCE-MRI) technique called GRASP-Pro (golden-angle radial sparse parallel imaging with imProved performance) through a joint sparsity and self-calibrating subspace constraint with automated selection of contrast phases.

Methods: GRASP-Pro reconstruction enforces a combination of an explicit low-rank subspace-constraint and a temporal sparsity constraint. The temporal basis used to construct the subspace is learned from an intermediate reconstruction step using the low-resolution portion of radial k-space, which eliminates the need for generating the basis using auxiliary data or a physical signal model. A convolutional neural network was trained to generate the contrast enhancement curve in the artery, from which clinically relevant contrast phases are automatically selected for evaluation. The performance of GRASP-Pro was demonstrated for high spatiotemporal resolution DCE-MRI of the prostate and was compared against standard GRASP in terms of overall image quality, image sharpness, and residual streaks and/or noise level.

Results: Compared to GRASP, GRASP-Pro reconstructed dynamic images with enhanced sharpness, less residual streaks and/or noise, and finer delineation of the prostate without prolonging reconstruction time. The image quality improvement reached statistical significance (P < 0.05) in all the assessment categories. The neural network successfully generated contrast enhancement curves in the artery, and corresponding peak enhancement indexes correlated well with that from the manual selection.

Conclusion: GRASP-Pro is a promising method for rapid and continuous DCE-MRI. It enables superior reconstruction performance over standard GRASP and allows reliable generation of artery enhancement curve to guide the selection of desired contrast phases for improving the efficiency of GRASP MRI workflow.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Feng L, Wen Q, Huang C, Tong A, Liu F, Chandarana H. GRASP-Pro: imProving GRASP DCE-MRI through self-calibrating subspace-modeling and contrast phase automation. Magn Reson Med. 2020;83(1):94-108. doi:10.1002/mrm.27903
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
Magnetic Resonance in Medicine
Rights
Publisher Policy
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}