Towards the use of a smartphone imaging-based tool for point-of-care detection of asymptomatic low-density malaria parasitaemia

dc.contributor.authorColbert, Ashlee J.
dc.contributor.authorCo, Katrina
dc.contributor.authorLima‑Cooper, Giselle
dc.contributor.authorLee, Dong Hoon
dc.contributor.authorClayton, Katherine N.
dc.contributor.authorWereley, Steven T.
dc.contributor.authorJohn, Chandy C.
dc.contributor.authorLinnes, Jacqueline C.
dc.contributor.authorKinzer‑Ursem, Tamara L.
dc.contributor.departmentPediatrics, School of Medicineen_US
dc.date.accessioned2023-03-09T15:05:12Z
dc.date.available2023-03-09T15:05:12Z
dc.date.issued2021-09-25
dc.description.abstractBackground: Globally, there are over 200 million cases of malaria annually and over 400,000 deaths. Early and accurate detection of low-density parasitaemia and asymptomatic individuals is key to achieving the World Health Organization (WHO) 2030 sustainable development goals of reducing malaria-related deaths by 90% and eradication in 35 countries. Current rapid diagnostic tests are neither sensitive nor specific enough to detect the low parasite concentrations in the blood of asymptomatic individuals. Methods: Here, an imaging-based sensing technique, particle diffusometry (PD), is combined with loop mediated isothermal amplification (LAMP) on a smartphone-enabled device to detect low levels of parasitaemia often associated with asymptomatic malaria. After amplification, PD quantifies the Brownian motion of fluorescent nanoparticles in the solution during a 30 s video taken on the phone. The resulting diffusion coefficient is used to detect the presence of Plasmodium DNA amplicons. The coefficients of known negative samples are compared to positive samples using a one-way ANOVA post-hoc Dunnett's test for confirmation of amplification. Results: As few as 3 parasite/µL of blood was detectable in 45 min without DNA extraction. Plasmodium falciparum parasites were detected from asymptomatic individuals' whole blood samples with 89% sensitivity and 100% specificity when compared to quantitative polymerase chain reaction (qPCR). Conclusions: PD-LAMP is of value for the detection of low density parasitaemia especially in areas where trained personnel may be scarce. The demonstration of this smartphone biosensor paired with the sensitivity of LAMP provides a proof of concept to achieve widespread asymptomatic malaria testing at the point of care.en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.citationColbert AJ, Co K, Lima-Cooper G, et al. Towards the use of a smartphone imaging-based tool for point-of-care detection of asymptomatic low-density malaria parasitaemia. Malar J. 2021;20(1):380. Published 2021 Sep 25. doi:10.1186/s12936-021-03894-wen_US
dc.identifier.urihttps://hdl.handle.net/1805/31761
dc.language.isoen_USen_US
dc.publisherBMCen_US
dc.relation.isversionof10.1186/s12936-021-03894-wen_US
dc.relation.journalMalaria Journalen_US
dc.rightsAttribution 4.0 International*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/*
dc.sourcePMCen_US
dc.subjectMalariaen_US
dc.subjectParticle-diffusometryen_US
dc.subjectNucleic-acid based testsen_US
dc.subjectSmartphone-detectionen_US
dc.subjectLAMP-assayen_US
dc.titleTowards the use of a smartphone imaging-based tool for point-of-care detection of asymptomatic low-density malaria parasitaemiaen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
12936_2021_Article_3894.pdf
Size:
1.67 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.99 KB
Format:
Item-specific license agreed upon to submission
Description: