Maximal torque and torque-twist rate were determined for straight lengths of wire. The effect of wire length on each was studied. Anti-rotational torque was determined for cuspid retraction springs, and finally, distal root torque produced by cuspid retraction springs and anterior retraction springs was determined.
Torque watch gauges in combination with a Bunsen burner assembly and self-devised deflection scale were used to determine the data.
Maximal torque was shown to be independent of wire length, and closely approximated the theoretically determined values, while twist was shown to be proportional to wire length.
Pre-stressing followed by stress-relief produced the highest maximal torque values, while wires which received neither type of treatment exhibited the lowest values.
Torque-twist rate was determined for a number of wires commonly used in orthodontics, both experimentally and theoretically.
It was shown that anti-rotational torque produced by cuspid retraction springs used in the Segmented Arch Technique was not only capable of preventing cuspid rotation during distal tooth movement, but could actually move the distal of the cuspid labially on the basis of theoretical mechanics.
Initial activation of anterior retraction springs used in the Segmented Arch Technique produces added distal root torque for the appliance when activated for distal tooth movement, while it appears that both cuspid and anterior retraction springs produce sufficient distal root torque to prevent forward root displacement.