COMPARISON OF I-123 THYROID UPTAKE PHANTOM VALUES: GAMMA CAM-ERA VERSUS UPTAKE PROBE
Date
Authors
Language
Embargo Lift Date
Department
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Abstract
The purpose of this study was to investigate whether gamma cameras and uptake probes give comparable uptake results using 123I capsules. Methods: During a month-long period, five standard 123I thyroid dose cap-sules (156-163 μCi) were placed into a neck phantom, and positioned 10 centimeters away from the uptake probes and pinhole collimators attached to two gamma cameras. Each capsule was counted sixty seconds in three trials per capsule, along with a sixty second background count. The counts measured by both uptake probes and gamma cameras were compared. Re-sults: Comparison between the gamma camera uptake values and the probe system uptake values were done. The Argus gamma camera had higher counts than the Transcam 80% of the time. The probe systems had higher counts than the gamma cameras 100% of the time. Conclusion: Probe A and Probe B had significant higher counts than the Transcam and the Argus camera.