    Identifying and troubleshooting low-frequency artifacts mimicking atrial flutter caused by deep brain stimulator
    (Elsevier, 2021-07-22) Stelzer, Mitchell A.; Padanilam, Mathew S.; Miller, Jennifer J.; Foreman, Jason; Golden, Keith; Ahmed, Asim S.; Medicine, School of Medicine
    Electromagnetic interference from deep brain stimulators (DBS) is known to cause high-frequency artifact on some forms of cardiac telemetry.1 Here, we (1) present a unique case in which a DBS caused a low-frequency artifact mimicking typical atrial flutter leading to referral to electrophysiology; (2) demonstrate a method to confirm this; and (3) provide recommendations of how to minimize artifact and work around this interaction in future rhythm monitoring attempts.