Effect of Firing Rate on the Performance of Shock Wave Lithotriptors

Date
2008
Language
American English
Embargo Lift Date
Committee Members
Degree
Degree Year
Department
Grantor
Journal Title
Journal ISSN
Volume Title
Found At
Wiley
Can't use the file because of accessibility barriers? Contact us with the title of the item, permanent link, and specifics of your accommodation need.
Abstract

Objective: To determine the mechanism that underlies the effect of shock wave (SW) rate on the performance of clinical lithotripters.

Materials and methods: The effect of firing rate on the pressure characteristics of SWs was assessed using a fibre-optic probe hydrophone (FOPH 500, RP Acoustics, Leutenbach, Germany). Shock waves were fired at slow (5-27 SW/min) and fast (100-120 SW/min) rates using a conventional high-pressure lithotriptor (DoLi-50, Dornier MedTech America, Inc., Kennesaw, GA, USA), and a new low-pressure lithotriptor (XX-ES, Xi Xin Medical Instruments Co. Ltd, Suzhou, PRC). A digital camcorder (HDR-HC3, Sony, Japan) was used to record cavitation fields, and an ultrafast multiframe high-speed camera (Imacon 200, DRS Data & Imaging Systems, Inc., Oakland, NJ, USA) was used to follow the evolution of bubbles throughout the cavitation cycle.

Results: Firing rate had little effect on the leading positive-pressure phase of the SWs with the DoLi lithotriptor. A slight reduction ( approximately 7%) of peak positive pressure (P+) was detected only in the very dense cavitation fields (approximately 1000 bubbles/cm(3)) generated at the fastest firing rate (120 SW/min) in nondegassed water. The negative pressure of the SWs, on the other hand, was dramatically affected by firing rate. At 120 SW/min the peak negative pressure was reduced by approximately 84%, the duration and area of the negative pressure component was reduced by approximately 80% and approximately 98%, respectively, and the energy density of negative pressure was reduced by >99%. Whereas cavitation bubbles proliferated at fast firing rates, HS-camera images showed the bubbles that persisted between SWs were very small (<10 microm). Similar results were obtained with the XX-ES lithotriptor but only after recognizing a rate-dependent charging artefact with that machine.

Conclusion: Increasing the firing rate of a lithotriptor can dramatically reduce the negative pressure component of the SWs, while the positive pressure remains virtually unaffected. Cavitation increases as the firing rate is increased but as the bubbles collapse, they break into numerous microbubbles that, because of their very small size, do not pose a barrier to the leading positive pressure of the next SW. These findings begin to explain why stone breakage in SWL becomes less efficient as the firing rate is increased.

Description
item.page.description.tableofcontents
item.page.relation.haspart
Cite As
Pishchalnikov YA, McAteer JA, Williams JC Jr. Effect of firing rate on the performance of shock wave lithotriptors. BJU Int. 2008;102(11):1681-1686. doi:10.1111/j.1464-410X.2008.07896.x
ISSN
Publisher
Series/Report
Sponsorship
Major
Extent
Identifier
Relation
Journal
BJU International
Source
PMC
Alternative Title
Type
Article
Number
Volume
Conference Dates
Conference Host
Conference Location
Conference Name
Conference Panel
Conference Secretariat Location
Version
Author's manuscript
Full Text Available at
This item is under embargo {{howLong}}