Li, GuangyanConnors, Bret A.Schaefer, Ray B.Gallagher, John J.Evan, Andrew P.2019-06-242019-06-242017-11Li, G., Connors, B. A., Schaefer, R. B., Gallagher, J. J., & Evan, A. P. (2017). Evaluation of an experimental electrohydraulic discharge device for extracorporeal shock wave lithotripsy: Pressure field of sparker array. The Journal of the Acoustical Society of America, 142(5), 3147. doi:10.1121/1.5010901https://hdl.handle.net/1805/19664In this paper, an extracorporeal shock wave source composed of small ellipsoidal sparker units is described. The sparker units were arranged in an array designed to produce a coherent shock wave of sufficient strength to fracture kidney stones. The objective of this paper was to measure the acoustical output of this array of 18 individual sparker units and compare this array to commercial lithotripters. Representative waveforms acquired with a fiber-optic probe hydrophone at the geometric focus of the sparker array indicated that the sparker array produces a shock wave (P+ ∼40-47 MPa, P- ∼2.5-5.0 MPa) similar to shock waves produced by a Dornier HM-3 or Dornier Compact S. The sparker array's pressure field map also appeared similar to the measurements from a HM-3 and Compact S. Compared to the HM-3, the electrohydraulic technology of the sparker array produced a more consistent SW pulse (shot-to-shot positive pressure value standard deviation of ±4.7 MPa vs ±3.3 MPa).en-USPublisher PolicyExtracorporeal shock wave sourceEllipsoidal sparker unitsKidney stonesCommercial lithotriptersDornier HM-3Dornier Compact SElectrohydraulic technologySparker arrayConsistent SW pulseAcoustic measurementsEvaluation of an experimental electrohydraulic discharge device for extracorporeal shock wave lithotripsy: Pressure field of sparker arrayArticle